1 - Qvision, Hospital Vithas Virgen del Mar, Almería, España
2 - Complejo Hospitalario de Torrecárdenas, Almería, España
Luminance is a physical magnitude that determines the amount of light that reflects or emits a stimulus. Do not confuse it with Luminosity, which is the perception of luminous intensity of the visual system. There are different ways to quantify the luminance contrast between a stimulus and its background. When there are only two levels of luminance, such as the luminance of the optotype (Lo) and the background (Lf) in a variable contrast optotype, the following equation is used to characterize the contrast (Figure 1A)1.
This equation has been written assuming that the background luminance is greater than that of the optotype. In case of positive optotypes (dark background, clear optotype) the numerator of the previous equation is inverted. Since the equation is based on a ratio of luminances, the resulting contrast (C) does not have any physical magnitude that goes with it. When more than two involved luminances appear in the test, such as in a sinusoidal network, we use Michelson’s contrast formula (Figure 1B)2:
These two equations, used to calculate the contrast, are the basis of all clinical tests for the evaluation of contrast sensitivity. The contrast is a physically measurable magnitude that corresponds to the information coming from the stimulus. When we present a variable contrast stimulus to an observer, decreasing its magnitude, there will come a time that it cannot be detected. We say then that we have reached the Contrast Threshold.
The threshold of contrast can be defined as the minimum amount of contrast needed to have the stimulus on a uniform background to be detected. If the contrast is lower than our threshold value, we will not be able to detect it. Then appears an ability of the observer that is related to its ability to detect an object on a background. This ability is called Contrast Sensitivity (CS). The contrast (physical magnitude, C) perceptible by an observer will reach a threshold value whose inverse will give rise to contrast sensitivity (visual ability, CS = 1/C).
Figure 1. A) Variable contrast optotype where Lo represents the luminance of the optotype and Lf that of the background. B) Variable contrast sinusoidal network where Lmax represents the maximum luminance of the network and Lmin the minimum. A cycle comprises a pair of maximum and minimum luminance lines. The frequency of the network is determined based on the number of cycles per unit of degree.
We already talked in the previous section that we can use sinusoidal or variable contrast optotypes to assess contrast sensitivity. However, the differences between the two visual stimuli mean that the information obtained with each of them is not interchangeable, although it is related in some way3,4.
Sinusoidal networks are the simplest stimulus pattern for measuring contrast sensitivity. As we can see in Figure 1B there is a degraded transition between each maximum and minimum luminance. This combination is called a cycle and will be used to describe the contrast sensitivity test. The more cycles there are within a degree we will say that the network has a greater spatial frequency and vice versa. In contrast sensitivity tests composed of sinusoidal networks, the so-called Contrast Sensitivity Function (CSF) is analyzed. To determine this curve, the sensitivity (1/C) is calculated with the increase of the spatial frequency. The result is a curve where the evaluated spatial frequencies are represented on the abscissa axis and the sensitivity for each of these frequencies on the ordered axis.
The sensitivity obtained by increasing the frequency of a sinusoidal network is correlated with the sensitivity obtained by decreasing the size of an optotype. Although it is true that this correlation is not linear5, the sensitivity measured with high frequency networks maintains a better agreement with the sensitivity measured with optotypes of small size or less detail and vice versa. Figure 2 describes an analogy between the CSF measured with sinusoidal networks and the agreement with optotypes of variable size. The curve would represent the CSF in such a way that contrasts above the curve will not be perceived by the patient as opposed to contrasts below the curve. We also see that as the frequency increases and the size of the optotype decreases so does the contrast sensitivity in the normal patient. A metric related to this curve would be the visual or inverse acuity of the smallest optotype detail that a patient is able to solve.
Both high- and low-contrast visual acuity would represent only one point of the curve, which shows the loss of information about total visual performance when we only measure visual acuity and not CSF.
The CSF measurement tests in multifocal intraocular lens studies do not measure the CSF completely but select 4 or 5 spatial frequencies. For example, the VCTS6,7 and the FACT8,9, two of the most used CSF measurement tests incorporate 5 spatial frequencies (1.5, 3, 6, 12 and 18), with the separation between frequencies being one octave. The CSV1000-E and the ClinicCSF10, on the other hand, only evaluate 4 channels (3, 6, 12 and 18) separated by an octave11 and inclining the sinusoidal networks ±15º to subject the patient to a multiple choice task with 3 possible answers (vertical, tilted to the right and tilted to the left)2.
Figure 2. The contrast sensitivity function represents the contrast detected by a patient based on their spatial frequency. The spatial frequency with sinusoidal networks correlates non-linearly with the size of the optotype. For a better understanding, the figure shows an optotype decreasing its size to describe the concept of spatial frequency instead of using a sinusoidal network with higher spatial frequency.
CSF is a common measure in studies of visual performance with Multifocal Intraocular Lenses (MIOLs). Marques et al12 did not find significant differences for 4 spatial frequencies of 3, 6, 12 and 18 cpg compared to FineVision and AT Lisa Tri. Ruiz-Mesa et al13 also did not find differences between FineVision and Symfony, including the frequency of 1.5 cpg and reporting that the low spatial frequencies of 1.5, 3 and 6 were within the normal range of patients not implanted with MIOLs. Jonker et al14 reported significant differences in favor of a bifocal Acrysoft Restor +3 lens versus a Finevision lens but exclusively within the frequency of 6 cpg. Cochener B et al6 also did not find significant differences between a bifocal lens Tecnis ZMB00 and a Finevision for the 4 frequencies evaluated of 1.5, 3, 6 and 12 cpg. Bilbao-Calabuig et al15 also found no significant differences between a blended-vision with two bifocals of +2.50 D and +3.00 D and the FineVision lens for no spatial frequency. Pedrotti et al16 also did not report significant differences between a Tecnis ZCB00 monofocal intraocular lens and the Symfony lens for any of the spatial frequencies evaluated.
In our opinion, as evidenced by the results of the aforementioned studies, the measurement of the CSF for 4 spatial frequencies could be replaced by other metrics that provide more information on visual performance with MIOLs at multiple distances since the CSF is usually measured only at a distance (far) and on this it has not been found different affectation for the 4 spatial frequencies in comparative between MIOLs.
Blur curves represent the most complete metric today to know the patient's visual performance at multiple distances. These types of curves have been carried out since the 1990s with the purpose of knowing the extension of the depth of field in a multifocal procedure and the vision achieved by the patient at multiple distances17.
To know what the vision of the patient is at multiple distances we could use a visual acuity test that was changing scale depending on the presentation distance. However, this would be a very laborious process, which would take us a long time, and not without errors due to not finding us at the exact evaluation distance, especially when we evaluate the near vision range in which a few centimeters can give us a significant variation in the patient's vision with a MIOL (Figure 3).
Figure 3. Correlation between the distance between the presentation of the test and the lens used to simulate the distance of presentation.
To expedite this procedure and minimize possible errors in the distance of presentation we can use a remote distance test (~ 6 m) and introduce different positive and negative lenses that vary the vergence of light in the same way as if we varied the distance of presentation of the test. To calculate the correspondence between distance and diopters we just have to calculate the inverse of the distance in meters. For example, to simulate a distance of 40 cm we would have to place a power lens equal to -2.5 D (1/-0.4 m).
The correlation between dioptric jumps and distance is not linear as shown in Figure 3. This means that a small variation in the near distance presentation may correspond to an important dioptric jump. As can be seen in the enlarged region of near in Figure 3. A variation in the presentation distance of 25 cm to 30 cm corresponds to a dioptric jump of more than 0.5 D from a blur lens of -4.00 D to a higher than -3.50 D while a dioptric jump of 0.50 D in far distance, for example from -1.0 D to -0.5 D, corresponds to a distance variation of 100 cm. Due to this important variation of visual acuity with a small distance error, it is advisable not to measure the near-distance blur curves and always do it at a distant distance18. We already know that it is recommended to measure the blur curves for distance, the greater the distance the better, and if we have no possibility of having a cabinet with a distance of 4 meters or more we can carry out the procedure at 2 m as long as we place a +0.50 D lens during the test that corrects the proximal distance.
The next question we can ask ourselves is, what blur lenses should I use? The answer to this question will depend on the range of vision we want to evaluate or in other words, the working distance for close of the patient. Bearing in mind this consideration, it is true that in a range of +1.00 D to -4.00 D we cover the patient's vision from infinite to 25 cm, as shown in Figure 3. In particular cases this range may vary, for example with intraocular lenses of extended depth of focus (+1.50 D to -2.50 D)19. In addition to the range of vision we must also decide the dioptric jumps that we are going to select; in a standardized way we will always carry out the curves in steps of 0.50 D20.
When we carry out the procedure with optotype charts in patients with intraocular lenses (pseudophakic) it is preferable to start the procedure with negative lenses (-4.00 D) since we will go from a worse vision to a better vision, and in this way we will avoid the problem of memorization of the letters in case we do not have a computerized test in which the optotypes vary randomly with each patient’s response. In addition, in bifocal intraocular lenses in which visual acuity decreases in intermediate vision it would be advisable to change the static optotype chart from -2.0 D in order to avoid memorizing the letters with the consequent over-estimation in intermediate vision.
Another alternative proposed by Gupta et al to avoid the learning effect is to randomize the reading order of the letters, randomize the order of the defocus lenses or both21. This last criterion of randomizing the defocus lens is the one currently used during many clinical trials with optotype charts. However, the aforementioned drawbacks disappear when we use a random optotype test such as the Multifocal Lens Analyzer22.
Blur curves of visual acuity are widely used in clinical practice. However, as we have already seen, visual acuity only represents one point in the CSF and patients with the same visual acuity can have a very different visual quality that can be differentiated through a contrast sensitivity test23. In 2016, we started the investigations with a new automated tool designed and programmed by Rodríguez-Vallejo M. for measuring the blur curves that allows not only to evaluate the visual acuity blur curve but also the contrast sensitivity blur curve from the contrast variation of a static-sized optotype22. This tool called Multifocal Lens Analyzer consists of an automated test reproduced through an iPad which has proven to be a useful and capable instrument for measuring visual performance10,24–26.
The system has important advantages with respect to conventional tests for measuring defocus curves, including:
Since the procedure is fully automated, the clinician only has to click on the button that matches the response given by the subject, with the application controlling whether to increase or decrease the size/contrast of the optotype, move to the next Defocus level through a lens shift alert message, or end the test. The measurement protocol with the Multifocal Lens Analyzer involves the following procedure:
Figure 4. Monocular defocus curve of visual acuity in a patient implanted with a low-addition trifocal intraocular lens.
The following is an example of a clinical case in which the contrast sensitivity blur curves more accurately represent the patient's vision quality than the visual acuity blur curves. Figure 4 depicts a blur curve of monocular visual acuity in a patient implanted with a low-addition trifocal intraocular lens (+3.00 D in the IOL plane, +2.00 D in the glasses plane).
Despite being taken with the best subjective correction we can verify that the visual acuity for far (0 D) is 0 logMAR (20/20). In addition, the visual acuity is maintained at 0.1 logMAR (20/25) to the blur lens of -4.00 D, which means a visual acuity of 20/25 up to a distance of 25 cm, something really strange in a low-addition trifocal lens. We could say that the visual acuity blur curve is really good, but does that mean that the quality of vision is as good as the visual acuity blur curve represents?
Figure 5. In the upper part, an image of Scheimpflug is shown, where we can see that the lens is not only off-centered but inclined. The lower image shows a slit lamp capture that reveals the decentration of the lens which, together with the inclination, causes a lower deformation of the pupil.
Figure 5 demonstrates that there is an anomaly whereby a flatter visual acuity blur curve is obtained. This anomaly consists of a decentralization and inclination of the lens, that induces an increase in the depth of focus, such that a good visual acuity is achieved higher than expected in intermediate vision and beyond -2.0 D.
However, this improvement in visual acuity does not imply an improvement in visual quality as shown in the contrast sensitivity blur curve of Figure 6.
Figure 6. Contrast sensitivity blurring measured with an optotype of size corresponding to a sharpness visual of 0.3 logMAR.
In this case we see a displacement of the -0.5 D curve that has not been detected by the visual acuity blur curve. In addition, in the contrast sensitivity blur curve a steeper slope of the visual quality increase from +1.00 D to -0.50 D is shown, while in the visual acuity blur curve only a variation of 1 line of sharpness was detected visual.
In summary, considering that the average contrast sensitivity of a series of patients implanted with a trifocal intraocular lens is usually about 0.8 logSC in distance vision and 0.5 logSC in near vision for an optotype size of 0.3 logMAR27, we can conclude that the patient has a considerable loss of visual quality in distance that is not visible in the visual acuity blur curves and that their depth of field in terms of contrast sensitivity is much smaller than that shown by the visual acuity blur curve.